Instruction Sheet
TROLL
®
9500 S2XP Restrictor Installation and Calibration
Part number 0085490, Rental R79420
Overview
The S2XP restrictor is shipped in two pieces. The
restrictor must be installed in separate steps to allow for
sensor calibration.
It is recommended you follow all steps in this instruction
sheet to properly set up your TROLL 9500 S2XP
system.
Contents
Figure 1.1 Contents of the shipping container
1 TROLL 9500 Instrument
2 S2XP restrictor, part A
3 Quick Cal solution
4 S2XP restrictor, part B
5 Sodium sulfite solution
6 Accessories kit
7 RDO sensor; TROLL Com
8 Bubbler (not shown)
Units may contain other equipment.
pH, ORP, and Conductivity Calibration
1. Detach the Cal Cup from the TROLL 9500. The Cal
Cup contains a small amount of liquid to hydrate the
pH sensor during shipment. Discard this liquid.
2. Thoroughly rinse the sensors with clean water.
3. Fill the Cal Cup to the fill line with Quick Cal solution.
Reinstall the Cal Cup on the TROLL 9500. See
Figure 1.2. Do not over-tighten.
Figure 1.2 Calibration using Quick Cal in the Cal Cup
4. In Win-Situ 4, click on Parameters in the Navigation
tree.
5. Click Quick Cal. Select pH, ORP, and Conductivity.
6. Click Start. You must hold the TROLL 9500 vertically
for proper calibration.
7. Allow the calibration to run until indicators reach
Stable. Check calibration coefficients to ensure a
proper calibration.
Information subject to change without notice. In-Situ, In-Situ logo, Baro Merge, BaroTROLL, HERMIT, iSitu, Pocket-Situ, RDO,
RuggedCable, RuggedReader, TROLL, and Win-Situ are trademarks or registered trademarks of In-Situ Inc.
©
2013. All rights reserved.
08/2013 | Rev. 005 | 0085492
Vue de la page 0
1 2 3 4 5

Résumé du contenu

Page 1 - Instruction Sheet

Instruction SheetTROLL®9500 S2XP Restrictor Installation and CalibrationPart number 0085490, Rental R79420OverviewThe S2XP restrictor is shipped in tw

Page 2

Instruction SheetpH slope Between -50 to -66 mV/pHpH offset 350 - 450 mVORP offset Between -20 and +20 mVK Cell Low 0.30 - 0.39 cm-1K Cell High 4.4 -

Page 3

Instruction Sheet10. Refresh Win-Situ 4 to ensure all sensors are correctlyattached to the TROLL 9500. pH, ORP, Conductivity,and Rugged Dissolved Oxyg

Page 4

Instruction Sheet5. In Win-Situ 4, click Run in the Zero Oxygen Pointscreen. Follow the on-screen instructions to completethe 0% calibration point.6.

Page 5

Instruction SheetFigure 1.14 Removing the pH/ORP sensor5. Remove the conductivity sensor using the removaltool. Note the sensor's port location (

Commentaires sur ces manuels

Pas de commentaire