1WIN-SITU 4.0 USER’S GUIDE 0031510 rev. 003 11/04Contents1 INTRODUCTION ... 6O
10 0031510 rev. 003 11/04The application window appears. A new, empty site called “Home” isdisplayed at the top of the Navigation tree. The Connecti
100rev. 003 11/04APPENDIX B: TEST MEASUREMENT SCHEDULESFirst Six Seconds of TestIn a true logarithmic sampling scheme, the interval between points wo
101rev. 003 11/04TIP: The storageinterval (A) mustbe at least twiceas long as the rapid interval(B) times the number ofMeasurements foraveraging (C).
102WIN-SITU 4.0 USER’S GUIDErev. 003 11/04APPENDIX C: THELOG SCHEDULEFor reference, the entire log schedule out to the 218th data point is shownon th
103rev. 003 11/04N = Data Point Number TIME TO NEXT POINT ELAPSED TIME N hr min sec hdrd day hr min sec hdrd 1700030 000
104rev. 003 11/04N = Data Point Number TIME TO NEXT POINT ELAPSED TIME N hr min sec hdrd day hr min sec hdrd 5100190 000
105rev. 003 11/04N = Data Point Number TIME TO NEXT POINT ELAPSED TIME N hr min sec hdrd day hr min sec hdrd 84 0 0 13 3
106rev. 003 11/04N = Data Point Number TIME TO NEXT POINT ELAPSED TIME N hr min sec hdrd day hr min sec hdrd117 0 1 29 3
107rev. 003 11/04N = Data Point Number TIME TO NEXT POINT ELAPSED TIME N hr min sec hdrd day hr min sec hdrd151 0103220
108rev. 003 11/04185 1 14 35 70 0 20 58 48 38186 119 090 0221324 8187 1 23 41 80 0 23 32 24 98188 1283940 1 056 678189 1335460 1 2244618190 1392850 1
109WIN-SITU 4.0 USER’S GUIDErev. 003 11/04IndexAABEND, 50, 52ActiveSync, 7, 8, 68, 76, 91Add... buttonadd connection, 19add test, 46BBatteriesdisplay
11 0031510 rev. 003 11/04• Devices connected: Select “one” to speed up connection to asingle device, or “more than one” for a network.• Port: Select
110rev. 003 11/04FFind button, 11, 15, 21Firmware upgrade, 31Flow-Sense Wizard, 75–89accepting values, 87creating templates, 76, 91–92entering target
111rev. 003 11/04exiting, 14, 69installation, 8, 70–72Navigation tree, 18requirements, 7Rugged Reader PDA, 7starting, 10starting in Profiler mode, 27
12 0031510 rev. 003 11/043. The following will happen:• Win-Situ synchronizes the baud rate of the device to the host PC.• Win-Situ “finds” (connect
13 0031510 rev. 003 11/04STEP 5: SET THE CLOCKData collection schedules depend on the device clock. Set it as follows:1. With the device selected, c
14 0031510 rev. 003 11/04THE NEXT STEPThe next step depends on the connected device and how you intend touse it.• Quick-Cal the water-quality sensor
15 0031510 rev. 003 11/04YOUR NEXT SESSIONEach time you start Win-Situ or Pocket-Situ, the site and the connection (ifany) saved in your last sessio
16WIN-SITU 4.0 USER’S GUIDE 0031510 rev. 003 11/043 THE NAVIGATION TREEWin-SItu and Pocket-Situ use the familiar “navigation tree” to display youri
17 0031510 rev. 003 11/04SECTION 3: THE NAVIGATION TREEWIN-SITU 4 INTERFACE (DESKTOP OR LAPTOP PC)TIP: In any view,clicking the rightmouse buttondis
18 0031510 rev. 003 11/04SECTION 3: THE NAVIGATION TREEPOCKET-SITU INTERFACE (PDA)TIP: Use the scrollbars to view moreof the Navigationtree and Info
19 0031510 rev. 003 11/04SITE VIEWWhen a site is selected, connection(s) and current status are shown.Site ActionsAdd starts the Connection Wizard t
2 0031510 rev. 003 11/04CONTENTSWin-Situ 4 Interface (Desktop or Laptop PC) ...
20 0031510 rev. 003 11/04DATA FOLDER VIEWThe Data Folder provides a convenient way to view tests that have beenextracted from the device to the host
21 0031510 rev. 003 11/04SECTION 3: THE NAVIGATION TREECONNECTION VIEWWhen a connection is selected, its properties are displayed.Connection Actions
22 0031510 rev. 003 11/04DEVICE VIEWWith a device selected, the Information pane shows details about thedevice: the date it was manufactured and cal
23 0031510 rev. 003 11/04MULTIPLE PARAMETERS VIEWTo view a list of all parameters the device can measure, select theParameters node in the Navigatio
24 0031510 rev. 003 11/04SECTION 3: THE NAVIGATION TREESINGLE PARAMETER VIEWTo view more detail for an individual parameter—including the pressurese
25 0031510 rev. 003 11/04MULTIPLE TESTS VIEWTo view a summary of all tests in the device, select the Tests node in theNavigation tree. Refer to Sect
26 0031510 rev. 003 11/04SECTION 3: THE NAVIGATION TREEWIN-SITU MENUS AND TOOLBAR (DESKTOP PC)File Menu—Use this menu to exit Win-Situ. It also has
27WIN-SITU 4.0 USER’S GUIDE 0031510 rev. 003 11/04CHOOSING MEASUREMENT UNITS AND OTHERPREFERENCESPreferences can be set at any time, and are saved w
28 0031510 rev. 003 11/04To display the Preferences window, do one of the following:• In Win-Situ, select Preferences from the Options menu.• In Poc
29 0031510 rev. 003 11/04OTHER PREFERENCESTo change other aspects of Win-Situ/Pocket-Situ operation, first displaythe Preferences window as describe
3 0031510 rev. 003 11/04CONTENTS6 WATER-QUALITY SETUP ... 387 PROFILING ...
30 0031510 rev. 003 11/04TAKING “MANUAL” READINGSDevices collect data in “tests” but you can get a quick “manual” readingfrom any device apart from
31 0031510 rev. 003 11/04UPGRADING FIRMWARE AND FEATURESFIRMWARE UPGRADENew software releases may include a new version of firmware for the In-Situ
32WIN-SITU 4.0 USER’S GUIDE 0031510 rev. 003 11/045 PRESSURE SETUPIn-Situ’s MiniTROLL measures two parameters: pressure (water level) andtemperatur
33 0031510 rev. 003 11/041. With the device connected to a host PC and selected in the Navigationtree, click or tap Pressure in the Parameters list.
34 0031510 rev. 003 11/04EDITING THE PRESSURE PARAMETERThe Parameter Wizard can help you configure a pressure channel todisplay measurements just as
35 0031510 rev. 003 11/04DISPLAY MODESelect one of the following. Subsequent screenswill be tailored to your choice.• Pressure Head displays the raw
36 0031510 rev. 003 11/04LEVEL/DEPTH CONVERSION INFORMATIONThe conversion from pressure in pounds per square inch (psi) to depth orlevel in meters o
37 0031510 rev. 003 11/04LEVEL REFERENCEThis is a user-specified starting point for level display modes. Here aresome examples:• Surface: Calculate
38WIN-SITU 4.0 USER’S GUIDE 0031510 rev. 003 11/046 WATER-QUALITY SETUPIn-Situ’s Multi-Parameter TROLL 9000 can measure water-qualityparameters in
39WIN-SITU 4.0 USER’S GUIDE 0031510 rev. 003 11/047 PROFILINGThe Profiler delivers instant real-time readings on all available channels.Profiler re
4 0031510 rev. 003 11/04CONTENTSPocket-Situ Graphing Controls ...
40 0031510 rev. 003 11/04SECTION 7: PROFILINGSTARTING THE PROFILER1. Connect the instrument to a PC andestablish communication in Win-Situor Pocket-
41 0031510 rev. 003 11/04SECTION 7: PROFILING• Readings are updated approximately every 2 seconds (default), orwhatever interval you may have specif
42 0031510 rev. 003 11/04SECTION 7: PROFILINGGRAPHING PROFILER DATATo see a graph of any channel during the current Profiler session, do oneof the f
43 0031510 rev. 003 11/04PROFILING WATER DEPTH OR LEVELProfiler readings from a pressure channel are displayed as pressure heador depth according to
44 0031510 rev. 003 11/04SECTION 7: PROFILINGLOGGING PROFILER DATAProfiler data may be logged to the connected PC while in the “Meter” view.• SnapSh
45WIN-SITU 4.0 USER’S GUIDE 0031510 rev. 003 11/048 LOGGING DATACollecting a set of data with an In-Situ instrument is called “running a test.”Many
46 0031510 rev. 003 11/04SECTION 8: LOGGING DATAADDING A TEST TO A DEVICE: THE TEST WIZARD1. Click or tap Tests in the Navigation tree.2. Click or t
47 0031510 rev. 003 11/04TEST SETUP PARAMETERSA complete test definition has five parts. The choices available in eachscreen are tailored to reflect
48 0031510 rev. 003 11/043. Measurement ScheduleThe available schedules depend on your device’s firmware and theparameters selected for inclusion in
49 0031510 rev. 003 11/044. Measurement IntervalThe measurement interval specifies how often measurements will be takenduring this test. You may ent
5 0031510 rev. 003 11/04CONTENTSOutput ...
50 0031510 rev. 003 11/045. Start Mode5a Manual. A manual start test can be started at any time while the deviceis connected to a PC by pressing the
51 0031510 rev. 003 11/04OTHER TEST OPERATIONSThere are six basic test operations:• Add...—programs the instrument to collect data. See the first pa
52 0031510 rev. 003 11/04The operations available depend on the status of the test selected in theNavigation tree. A dfiferent symbol is displayed f
53 0031510 rev. 003 11/04STARTING A TESTA test set up for a Scheduled start will start by itself, assuming the deviceclock is correctly set. To star
54 0031510 rev. 003 11/04EXTRACTING LOGGED DATARetrieving, downloading, uploading, transferring—these terms aresometimes used interchangeably to mea
55 0031510 rev. 003 11/04SECTION 8: LOGGING DATAWin-Situ \ Data \ SN00120 2004-06-18 083000 drawdown.binDevice serial numberTest start date (yyyy
56 0031510 rev. 003 11/04CLONING A TESTThe Clone function lets you copy all the elements of an existing test—thetest name, selected channels, measur
57WIN-SITU 4.0 USER’S GUIDE 0031510 rev. 003 11/04Logged data from In-Situ instruments are called “tests.” Tests reside in theinstrument until you e
58 0031510 rev. 003 11/04SELECTING DATA IN THE DATA FOLDERThe Data Folder appears below the Home site in the Navigation tree. Afterdata files have b
59 0031510 rev. 003 11/042. The Data Folder has a node for data from various device types—forexample, miniTROLL Pro Data, TROLL 9000 Professional XP
6WIN-SITU 4.0 USER’S GUIDE 0031510 rev. 003 11/041 INTRODUCTIONOVERVIEW OF WIN-SITU 4Win-Situ™ 4 is easy-to-use software that connects your desktop
60 0031510 rev. 003 11/044. The serial number node contains the extracted tests and loggedProfiler data. Click to select the test you want to view.T
61 0031510 rev. 003 11/045. To view the test data below the header in Report view:• Use the vertical and horizontal scrollbars to scroll through the
62 0031510 rev. 003 11/04GRAPHING DATA6. With a test selected in the Data Folder, click or tap Graph to displaythe individual parameter data in grap
63 0031510 rev. 003 11/04Each graph can display one, two, or all test channels. To change theappearance and content of any graph, display the Graph
64 0031510 rev. 003 11/04• To show two channels on the graph: select Graph Settings on theGraph menu, select a Primary channel and a Secondary Chann
65 0031510 rev. 003 11/04POCKET-SITU GRAPHING CONTROLSOn the Pocket PC, all channels in the test are displayed in one graph.Time is shown on the X a
66 0031510 rev. 003 11/04SECTION 9: VIEWING TEST DATAChanging the Data DisplayThe View... option in the command bar launches wizards to help youchan
67 0031510 rev. 003 11/04SECTION 9: VIEWING TEST DATATRANSFERRING DATA TO EXCEL (DESKTOP PC)1. In Win-Situ, select Report view.2. On the File menu,
68WIN-SITU 4.0 USER’S GUIDE 0031510 rev. 003 11/0410 POCKET-SITUPocket-Situ provides instrument control in the field using a handheld PDA(Personal D
69 0031510 rev. 003 11/04• To select measurement units and other preferences, select the Homesite in the Navigation tree and tap Setup on the comman
7 0031510 rev. 003 11/04PROGRAM REQUIREMENTSWIN-SITUWin-Situ requires a desktop or laptop computer running Microsoft®Windows® 95 or later with a ser
70 0031510 rev. 003 11/04INSTALLATIONThe In-Situ synchronization utility PocketSync can automatically install orupdate Pocket-Situ on a connected Ru
71 0031510 rev. 003 11/043. Install Win-Situ 4 on your desktop PC from the Win-Situ/Pocket-SituCD.4. Install Pocket-Situ on the desktop PC from the
72 0031510 rev. 003 11/04• If Pocket-Situ is not installed on the PDA, as shown in the screenabove, or needs an update, you will be asked if you wan
73 0031510 rev. 003 11/042. Double-click on “MyPocket PC.”3. Browse to the folderwhere Pocket-Situ Datais stored. On manyPDAs this will be in afolde
74 0031510 rev. 003 11/04SECTION 10: POCKET-SITUTROUBLESHOOTING SOFTWARE CONNECTIONSDESKTOP COM PORT CONFLICTSWindows 95 and Windows NT do not suppo
75rev. 003 11/04WIN-SITU 4.0 USER’S GUIDE11 FLOW-SENSE WIZARDIn-Situ’s low-flow software can help to determine when a representativesample of formati
76SECTION 11: FLOW-SENSE WIZARDrev. 003 11/04• Total well depth• Depth to water level, referenced to a benchmark• Depth to top of screen and placemen
77SECTION 11: FLOW-SENSE WIZARDrev. 003 11/04USING THE FLOW-SENSE WIZARDSTART THE SOFTWARELaunch Win-Situ or Pocket-Situ.If you have not used the sof
78SECTION 11: FLOW-SENSE WIZARDrev. 003 11/04• If you’ve used the Flow-Sense Wizard before—or copied “template”files from a different PC—expand the F
79SECTION 11: FLOW-SENSE WIZARDrev. 003 11/04Verify Unit PreferencesBefore the Flow-Sense Wizard starts, you may wish to verify the currentunit selec
8 0031510 rev. 003 11/04SECTION 1: INTRODUCTIONINSTALLATIONWIN-SITUThe Win-Situ CD-ROM contains an automatic setup program to installWin-Situ to you
80SECTION 11: FLOW-SENSE WIZARDrev. 003 11/041. The first input screen provides for entry of specific details about theproject. This information will
81SECTION 11: FLOW-SENSE WIZARDrev. 003 11/042. In the next screen, enter information about the well. This informationwill be included in the output
82SECTION 11: FLOW-SENSE WIZARDrev. 003 11/043. In the next screen, enter details about the pump and tubing. Tubinginformation is used in later volum
83SECTION 11: FLOW-SENSE WIZARDrev. 003 11/044. The next screen requests pumping informationFinal pumping rate: Enter in milliliters per minute.Total
84SECTION 11: FLOW-SENSE WIZARDrev. 003 11/04Measurement interval—Auto Calculated Value: This is the software-calculated time (seconds) for one compl
85SECTION 11: FLOW-SENSE WIZARDrev. 003 11/04Select Channels &Stabilization5. The final screen displays the Low Flow parameters. Choose theparame
86SECTION 11: FLOW-SENSE WIZARDrev. 003 11/046. Ensure the pump is on, then click or tap Start to begin the stabilizationreadings.STABILIZATION READI
87SECTION 11: FLOW-SENSE WIZARDrev. 003 11/04The graph shows the change and stabilization ranges of all parameters.The central area of the graph (the
88SECTION 11: FLOW-SENSE WIZARDrev. 003 11/04SAVING THE LOW FLOW DATAClick Save when a suggested output file name is displayed. You maychange this na
89SECTION 11: FLOW-SENSE WIZARDrev. 003 11/04EXPORT TO EXCEL OPTIONTo automatically create an output report in Microsoft® Excel™ from Win-Situ1. In t
9WIN-SITU 4.0 USER’S GUIDE 0031510 rev. 003 11/042 GETTING STARTEDSTEP 1: CONNECT THEINSTRUMENT TO THE PC1. Attach a communication interfaceto the
90SECTION 11: FLOW-SENSE WIZARDrev. 003 11/04Sample OutputReport%ODGGHU39&>FP@>P@>P@3HUFHQWDJH>P/PLQ@&
91SECTION 11: FLOW-SENSE WIZARDrev. 003 11/04PRIMING THE FLOW-SENSE WIZARD IN THE OFFICETo reduce the need for input on a PDA in the field, you may w
92SECTION 11: FLOW-SENSE WIZARDrev. 003 11/04In the Field4. Connect the TROLL 9000 to the PDA and navigate to the Flow-SenseWizard/Flow-Sense Data br
93WIN-SITU 4.0 USER’S GUIDErev. 003 11/04The conversion from PSI units to meters or feet of water requires severalconversion factors. One of these is
94rev. 003 11/04DETERMINING DENSITYThe density of water in a well can be determined using an accurate tapemeasure and a pressure transducer. A change
95rev. 003 11/043. Take an electronic pressure reading, manually, from the transducerand record this measurement in PSI. This is measurement P1.4. R
96WIN-SITU 4.0 USER’S GUIDErev. 003 11/04APPENDIX B: TESTMEASUREMENT SCHEDULESLINEARIn Linear sampling, the selected channels are measured at the sam
97rev. 003 11/04EVENTThis is much like Linear sampling with one important difference. Allselected parameters are measured at your chosen Measurement
98rev. 003 11/04Default StorageIn addition to Delta points, another form of data is also stored. Consider asituation where there is virtually no chan
99rev. 003 11/04LOGLogarithmic (or simply “log”) sampling optimizes data collection speed andmemory usage for pump and slug tests. These tests charac