RDO sensor
For best results, perform a 2-point calibration at these points before going into the
field.
100% calibration point:
Fill bubbler cup with clean water. Turn on the pump, regulate with the pinch a.
valve. Insert the TROLL 9500 into the bubbler, resting the RDO adapter
plate on the rim of the cup. Allow 5-10 minutes for temperature and oxygen
equilibration in the water. Select RDO in the software Navigation tree, click
the Calibrate button, then click Run.
B1
When readings have stabilized, click b. Next. Continue with the Zero Oxygen
Point for a 2-point calibration.
0% calibration point:
Fill the orange-based cal cup with sodium sulfite a.
(Na
2
SO
3
) solution and slide it up around the RDO
sensor until the cap and thermistor are completely
submerged. Be sure that there are no air bubbles
on the foil.
Select b. Run in the Zero Oxygen Point screen. When
readings have stabilized, click Next, then Finish to
finalize the calibration.
Rinse c. very thoroughly after calibrating in sodium
sulfite.
The RDO sensor is now calibrated and ready to use.
Turbidity sensor:
The sensor has been factory-calibrated over its full range (0
to 2000 NTU). If field calibration is required, we recommend a zero point calibration.
Use deionized water and the clear flow cell. Make sure that the 3-way valve on the
inflow port is closed.
In the office: Assemble the site data6
The software wizard will prompt for the following information for the output report:
• Well diameter & depth • Pump model & type • Tubing type
• Screen length • Depth to water, top of screen & pump intake
• Tubing inner diameter and length (for software volume calculations)
Install batteries1
Install D-cells in the TROLL 9500a. : Remove the battery
compartment cover, insert batteries, replace cover, and tighten.
Install and charge the RuggedReader battery pack: b.
This process is illustrated in the RuggedReader Quick Start.
To calibrate 100% dissolved oxygen, install alkaline D-cells in the c.
portable air pump as shown on the diagram inside the lid.
Connect the cable2
Remove the red dust caps. Insert the TROLL 9500 connector a.
into the cable connector. Push, twist, and click to lock. The
“click” ensures that the cable is securely attached.
Connect the TROLL Com to the other end of the cable.b.
Install software on your desktop/laptop PC from the 3
In-Situ
®
software CD
Win-Situ
®
4
Microsoft
®
ActiveSync
®
: Click on Microsoft Tools, select ActiveSync 4.X.
Win-Situ Sync
TIP: Although Pocket-Situ 4 is pre-installed on the RuggedReader. You need
ActiveSync and Win-Situ Sync on the PC to transfer files between the computers.
In the office: Launch the software4
Connect the TROLL 9500 to the PC and launch Win-Situ 4. Click a. Next through 4
screens of the Connection Wizard to set up the COM port.
Select the Com Port, tap b. Find.
TIP: Calibration and low-flow setup can optionally be done on the RuggedReader using
Pocket-Situ 4.
In the office: Calibrate the water quality sensors5
pH, ORP, and conductivity sensors
Remove the hydration bottle from the pH/ORP sensor. a. Save it to
keep the sensor hydrated between uses.
Use the orange-base cal cup with cutout and indentation. Fill with b.
Quick Cal, slide it up around the sensors, including the RDO cable.
In Win-Situ, tap c. Parameters, tap QuickCal, follow the Wizard.
TIP: Detailed steps for calibration and low-flow software are in the MP TROLL 9500
Operators Manual on the software CD.
RDO cable outside cup
RDO sensor inside cup
Low-Flow with
RDO
®
Optical Dissolved Oxygen
Quick Start
Bottom Layer of Case
Multi-Parameter
TROLL
®
9500
with RDO, Conductivity,
Turbidity (optional),
& pH/ORP sensors—
Remove hydration
bottle from pH/ORP
sensor before use!
Quick Cal
Solution
6’
Cable
Air pump
Flow
cell
Cal cup
for Quick Cal
Bubbler
contains
batteries
Hardware fittings
for flow cell
Alkaline batteries
for TROLL 9500
Top Layer of Case
RuggedCable
TROLL
®
Com
TROLL
®
9500
Flow cell
RDO cable inside cup,
RDO sensor outside cup
12 V adapter
for RuggedReader PC
AC adapter
for RuggedReader
®
PC
RuggedReader
®
battery
Win-Situ
software CD
and Quick Start Guides
TROLL
®
Com
serial comm cable
USB cable
for RuggedReader PC
RuggedReader
®
handheld PC
TROLL 9500
maintenance kit
Low-Flow with RDO
Quick Start Guide
Low Flow Base Plate
(not pictured)
Vue de la page 0
1 2

Résumé du contenu

Page 1

RDO sensorFor best results, perform a 2-point calibration at these points before going into the field.100% calibration point: Fill bubbler cup with cle

Page 2 - Low-Flow with RDO

Information is subject to change without notice. In-Situ and the In-Situ logo, Win-Situ, TROLL, BaroTROLL, RuggedReader, RuggedCable, and RDO are trad

Commentaires sur ces manuels

Pas de commentaire