Information Sheet
Due to continuing product development this information is subject to change without notice. In-Situ, the In-Situ logo, and TROLL are trademarks or regis-
tered trademarks of In-Situ Inc. Copyright © 2002–2006 by In-Situ Inc. All rights reserved.
1 800 446 7488
(toll-free, US and Canada) or 970 498 1500 www.in-situ.com
Multi-Parameter
Water Quality
TROLL
®
Catalog No. 33250
0032825 rev. 003 3/06
QUICK CAL
SOLUTION
pH, ORP, & Conductivity values
at various temperatures
pH mV µS/cm
Temp. (± 0.02) (± 5) (± 40)
5
o
C 7.10 255 4910
10
o
C 7.04 247 5635
15
o
C 7.03 239 6390
20
o
C 7.02 231 7180
25
o
C 7.00 224 8000
30
o
C 6.98 217 8850
35
o
C 6.97 209 9735
Specific Conductivity Coefficient:
1.95% /
degree C (0-25
o
C)
2.06% /
degree C (15-35
o
C)
2.23% /
degree C (25-50
o
C)
APPLICATION
A calibration standard of an appropriate chemical composition to perform a one-
point calibration of all four MP TROLL Basic sensors (pH, ORP, Conductivity,
Polarographic Dissolved Oxygen) at the same time with a single solution. Four
250 mL containers, each sufficient for at least 2 Quick Cals.
Shelf Life, unopened 6 months
After opening 7-21 days
EXPECTED ACCURACIES
The following accuriacies may be expected when calibrating the MP TROLL
using Quick Cal solution and following instructions in the Win-Situ Quick Cal
Wizard and/or in the MP TROLL Operator's Manual:
GUIDELINES AND PRECAUTIONS
Store the solution cool and dark. Shake well before use. For best results,
calibrate at the temperature expected during field use
Do not store sensors in Quick Cal solution
For best results, use fresh solution for each Quick Calibration
Dispose of used solution by flushing down the drain with plenty of water
Solution is best if used before date on bottle, but may be usable beyond that
date if stored cool, dark, and tightly capped
Quick Cal Meets USGS
Parameter accuracy guidelines*
pH: at pH 7 ± 0.09 pH units
over range 0-14 pH ± 0.2 pH units
ORP ± 25 mV NA
DO (with stirrer) ± 0.25 mg/L
Conductivity:
Low Range (350,000 µS/cm) ± 3%
High Range (70 – 200,000 µS/cm) ± 3%
* Calibration criteria for continuous water-quality monitors, from USGS Water-Resources
Investigations Report 00-4252, p. 13