Instruction Sheet
SMARTROLL MP Low-Flow Kits and Accessories
Part Number 0094250 (Complete Low-Flow Kit), 0093480 (Fittings Kit), 0094400 (Flow Cell Kit)
Application
The Low-Flow Sampling equipment is used to direct the
flow of water past the sensors of the SMARTROLL MP
Instrument during purging and low-flow sample tests.
The seal between the instrument and the flow cell
prevents water oxygenation prior to analysis. The 3-
way intake valve can be used to divert flow for sample
collection. The anti-siphon check valve prevents back
flow at the outlet. The system accommodates 3/8" and
1/4" tubing, which is not included.
Low-Flow Kit Contents
The Low-Flow Kit contains all of the items in the Fittings
Kit and the Flow Cell Kit. The base plate is sold
seperately.
Fittings Kit Contents
l 3/8" NPT to 3/8" barbed adapter fittings (2)
l 3/8" NPT to 1/4" barbed adapter fittings (2)
l 3/8" cap plug (1)
l 3/8" and 1/4" check valve (1 each)
l 3/8" and 1/4" 3-way ball valve (1 each)
l Pipe thread tape
Flow Cell Kit Contents
l Flow Cell with removable end cap (installed with
O-ring and stainless steel sleeve)
l Ground spike
Base Plate Kit
l 12" x 12" PVC, with hardware
Set up
1. Select the NPT/adapter fitting that matches the tubing
size.
2. Apply tape to the adapter fittings and twist them into
the threaded holes on the flow cell cylinder. Tighten
until hand tight. Do not over tighten. Ensure that the
fittings do not protrude beyond the wall of the flow cell.
3. If you are using the 3-way valve, attach it to the intake
fitting (bottom) with a short piece of tubing.
4. If you are using the anti-siphon check valve, attach it
to the outlet fitting (top) with a short piece of tubing.
5. The arrow on the valve indicates the flow direction.
6. To provide stability, attach the flow cell to the base
plate with the screw that is provided. Alternately, you
can use the metal spike and secure it in the ground.
Information subject to change without notice. In-Situ, In-Situ logo, BaroMerge, BaroTROLL, HERMIT, iSitu, Pocket-Situ, RDO,
RuggedCable, RuggedReader, TROLL, and Win-Situ are trademarks or registered trademarks of In-Situ Inc.
©
2013. All rights reserved.
04/2013 | Rev. 001 | 0094252
Vue de la page 0
1 2

Résumé du contenu

Page 1 - Instruction Sheet

Instruction SheetSMARTROLL™ MP Low-Flow Kits and AccessoriesPart Number 0094250 (Complete Low-Flow Kit), 0093480 (Fittings Kit), 0094400 (Flow Cell Ki

Page 2

Instruction Sheet7. Calibrate the instrument as shown in the Operator'sManual and in the mobile Help.8. Remove the restrictor from the instrument