Information Sheet
Catalog No. 0053310
SHORT STRIPPED & TINNED
CABLE
with Male Twist-Lock Connector
Desiccant
Dust cap—Remove
before use
Strain relief
Twist-Lock
connector
A
B
C
Dust cap
Vent tube
and
secured to
cable with
strap
Outboard
desiccant
attached to
cable vent
tube
(over)
Aqua
TROLL
®
Accessories
for Aqua TROLL
®
200
Connects an Aqua TROLL's RuggedCable
or to a controller or logger for communica-
tion via:
Analog (4-20 mA) RS485 Modbus
SDI-12 RS232 Modbus (with a customer-supplied converter)
Male Twist-Lock connector mates with female Twist-Lock connector on the RuggedCable
Stripped & tinned wires on opposite end for wiring to PLC or logger
Removable, refillable outboard desiccant protects cable vent tube from condensation in
high-humidity environments
SPECIFICATIONS
Materials PVC, titanium, nylon, Viton
®
Dimensions 13.33 cm (5.25 in) long, 19 mm (0.75 in) O.D.
Environmental Rating IP67 when connected, not designed for submersion
Cable 0.9 m (3 ft) vented TPU (Thermoplastic Polyurethane)
Conductors 6 conductors, 24 AWG, polypropylene insulation
Desiccant pack Clear acrylic, nylon, HDPE membrane
Desiccant approx. 22 g silica gel, type 4
Max. recommended SDI-12 61 m (200 ft)
cable lengths Modbus 1259 m (4000 ft)
4-20 mA 1259 m (4000 ft)
CONNECTIONS
Refer to wiring diagrams on back of sheet. Trim back and insulate unused wires. The
shield should be wired to a chassis ground or earth ground.
Power Connections. The Red wire provides power for all Aqua TROLL modes. The 4-
20 mA current loop output can be continuous in Modbus or SDI-12 mode as long as
Modbus device register 49507 is set to 1.
Communications. The device automatically switches between Modbus and SDI-12
modes depending on which of the two interfaces has activity. Modbus and SDI-12
cannot be used at the same time — whichever one is currently in use will block commu-
nication on the other.
To communicate with the device using Win-Situ
®
5 software through a serial COM port:
Disconnect the RuggedCable from the short stripped and tinned cable connector. Attach
a TROLL Com to the RuggedCable. Connect the TROLL Com's DB9-pin connector to a
serial port on a PC running Win-Situ 5.
A
B
C