Instruction Sheet
RDO
®
Classic Sensor Cap Replacement Kit
Part Number 0084230
Kit Contents
l RDO Classic Sensor Cap
l O-rings (2)
l O-ring lubricant
l Lens cloth
l Instruction sheet
Replace the Sensor Cap
The sensor cap has a 1-year typical life, and a 15-month
total life, after the instrument takes its first reading.
Install the cap by the date printed on the package.
Replacement caps are available from In-Situ Inc., or
from your authorized In-Situ distributor.
Do not allow moisture, including
atmospheric humidity, inside the cap.
Keep the cap in its sealed package until
you are ready to install it. Install promptly.
Make sure that O-ring grooves are dry and
that the O-rings are not rolled or pinched
inside the cap.
1. Remove the restrictor from the instrument.
2. Without twisting, pull the used sensor cap off of the
sensor.
3. Remove the existing O-rings from the sensor.
4. Use a lint-free cloth to remove any moisture from the
sensor body.
Make sure that the O-ring grooves are dry.
Avoid touching or cleaning the lens with
anything other than the supplied lens cloth.
5. Use your finger to apply a thin layer of lubricant
around the O-ring grooves. Place the O-rings on the
sensor.
Do not transfer lubricant to the lens or
sensor pins.
6. Clean the lens on the sensor with the cloth provided
in the kit and allow the lens to dry thoroughly.
Inspect for scratches or dirt.
7. Remove the new cap from its sealed package.
8. Align the arrow on the cap with the index mark on
the sensor and press it firmly, without twisting, until
it seals over the probe body. Make sure that the O-
rings are not pinched or rolled between the cap and
sensor.
9. Replace the restrictor.
10. Perform a 2-point calibration, if desired, as
described in the instrument manual.
1 Lightly coat with lubricant. Install 2 O-rings.
2 Clean lens and allow to dry.
3 Replace cap in a dry environment.
Information subject to change without notice. In-Situ, In-Situ logo, Baro Merge, BaroTROLL, HERMIT, iSitu, Pocket-Situ, RDO,
RuggedCable, RuggedReader, TROLL, and Win-Situ are trademarks or registered trademarks of In-Situ Inc.
©
2013. All rights reserved.
12/2013 | Rev. 004 | 0084232
Vue de la page 0
1

Résumé du contenu

Page 1 - Instruction Sheet

Instruction SheetRDO®Classic Sensor Cap Replacement KitPart Number 0084230Kit Contentsl RDO Classic Sensor Capl O-rings (2)l O-ring lubricantl Lens cl

Commentaires sur ces manuels

Pas de commentaire