Multi-Parameter TROLL 9500 OPERATOR’S MANUALJanuary 2009Multi-Parameter Water Quality TROLL®
30095110 rev. 007 01/09TROLL 9500 Operator’s Manual WHAT WE PROVIDEWARRANTY PROVISIONSIn-Situ Inc. warrants the Multi-Parameter TROLL 9500 for one y
93TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0910. When readings have stabilized, the screen will appear similar to the one below, and you hav
94TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09TIP: A warning box will appear if the slope is out of range (normal range is 0.9 to 1.1 with an
95TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SECTION 13: DISSOLVED OXYGEN—OPTICAL The values will be written to the sensor and you will be
96TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SECTION 13: DISSOLVED OXYGEN—OPTICALMAINTENANCE AND STORAGECleaning the Sensor Cap1. Leave the
97Multi-Parameter Water Quality TROLL®0095110 rev. 007 01/09TROLL 9500 Operator’s Manual 14 OXIDATION-REDUCTION POTENTIALFluid ORP (mV)Salt water
98TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09sensor removal toolPressure/Turbidity (or plug)Temperature Install pH/ORP sensor in port 1 onl
99TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0936789Indicators during Calibrations Status: NOT TESTED is displayed until you begin the calibr
10 0TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0911. The final screen shows the calculated sensor offset. This value should be 0 mV ± 20 mV.12
101TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09sensor removal toolUSAGE RECOMMENDATIONS AND CAUTIONSs /20READINGSVARYSLIGHTLYWITHTEMPER
104Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0915 AMMONIUMWHAT IS AMMONIUM ? Ammonium (NH4+) is the ion
4TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09CLEANING GUIDELINESPlease help us protect the health and safety of our employees by cleaning and
105TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SENSOR INSTALLATIONUnpack the ammonium sensor, hydrate it as above, and install in port 1, 2,
10 6TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09Specialized calibration kits are available for calibrating the ammonium sensor for low-range
10 7TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09 The Ammonium Calibration Wizard starts. A screen like this is displayed.8. Select the numb
108TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0916. The final screen of the Calibration Wizard shows the sensor slope and offset calculated dur
10 9TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SENSOR CARE AND HANDLINGSENSOR REMOVALPosition the yoke of the sensor removal tool at the poi
110Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0916 CHLORIDEWHAT IS CHLORIDE?Chloride (Cl–) is a highly s
111TROLL 9500 Operator’s Manual 0095110 rev. 007 01/092. Remove the sensor’s protective cap or storage bottle and set aside for future storage of t
112TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09CHLORIDE CALIBRATION PROCEDURE1. Rinse the Cal Cup and front end of the MP TROLL 9500 in clea
113TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0910. In the next screen, select Run to begin the stabilization. The display will continuously
114TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09The chloride sensor is now calibrated and ready to use.TIP: You can look at the calibration re
5Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SYSTEM DESCRIPTION The Multi-Parameter TROLL 9500 is a comp
115Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0917 NITRATEUnpolluted groundwater 0.05 ppm N Surface wat
116TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SENSOR PREPARATIONTo ensure optimum membrane response, the nitrate sensor should be thoroughly
117TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09s Single-point calibration. After the initial three-point bithermal cali-bration has establis
118TROLL 9500 Operator’s Manual 0095110 rev. 007 01/096. Select Nitrate in the Parameters list. The sensor serial number (SN) and recent calibrati
119TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0912. Remove the Cal Cup, discard the first solution, rinse the Cal Cup and the front end of the
12 0TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09POTENTIAL INTERFERENCESThe following table shows concentrations of possible interfering ions
121Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09WHAT IS TURBIDITY?Turbidity is an indirect measure of the
122TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09be detected. The plate is then pulled up until it is visible again. The average of the two dep
12 3TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0912 3 4Turbidity SensorDetector EmitterTemperature Install wiper in port 3Wiper brack-Q:A:What
12 4TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09Q:A:Why do I need to calibrate the turbidity sensor if it has been calibrated in the factory?
60095110 rev. 007 01/09TROLL 9500 Operator’s Manual CABLES Several basic cable types are used in the MP TROLL 9500 system:s 2UGGED#ABLEÛ405JACKE
12 5TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09Polymer-based standards are submicron, non-surface charged, solid spheres in matrixes of ultr
12 6TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09s 3INGLE0OINT#ALIBRATION#LICKNext and go to step 12.s 3TANDARD#ALIBRATIONDEFAULT8
12 7TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09131218 A screen similar to the one shown below is displayed.12. If you are performing perfor
12 8TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09RESETTING DEFAULT COEFFICIENTSThe sensor’s calibration may be reset back to factory defaults
12 9TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09Cleaning may be necessary if the optical windows of the sensor become visibly contaminated by
13 0Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09 Cable length Power supply feet meters req
131TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09 Order Parameter Unit 1 Pressure PSI 2 Temperature degrees Celsius 3 Barometric Pressure
132TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SECTION 19: SDI-12 OPERATIONs )FASENSORPORTISEMPTYTHEUNITWILLREPORTREADINGSWITHV
13 3TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SDI-12 V 1.3 COMMAND SETNAME COMMAND RESPONSE & COMMENTS!DDRESS1UERY A#2,&
13 4TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SDI-12 Data Re-corder (customer-supplied)MP TROLL 9500up to 300 ft (91 m) RuggedCable; see Ta
70095110 rev. 007 01/09TROLL 9500 Operator’s Manual CABLE DESICCANTSSmall Desiccant Clear cap of indicating silica gel desiccant seals the uphole en
13 5Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0920 LOW-FLOW MONITORINGGroundwater that has sat in a wel
13 6TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09START THE SOFTWAREConnect the TROLL 9500 to the PC or PDA. Launch Win-Situ 4 or Pocket-Situ 4
13 70095110 rev. 007 01/09TROLL 9500 Operator’s Manual SECTION 20: LOW -FLOW MONITORINGFLOW-SENSE WIZARD INPUT1. The first input screen provides fo
13 80095110 rev. 007 01/09TROLL 9500 Operator’s Manual 6. Measurement interval—Auto Calculated Value: This is the software-calculated time (seconds
13 90095110 rev. 007 01/09TROLL 9500 Operator’s Manual OUTPUT,OW&LOWlLESARESAVEDASlLESOFTYPEmOINAFOLDERNAMED&LOWSense Data u
14 0TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SECTION 20: LOW -FLOW MONITORINGSample of a Low-Flow output report using the Export to Excel
141Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0921 CARE & MAINTENANCEREPLACING BATTERIES The MP TROL
14 20095110 rev. 007 01/09TROLL 9500 Operator’s Manual REPLACEMENTIf the o-rings become damaged to the extent that no longer provide an effective se
14 3Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0922 TROUBLESHOOTINGQ:A:The MP TROLL 9500 rolled out of t
14 40095110 rev. 007 01/09TROLL 9500 Operator’s Manual TROUBLESHOOTING SENSORSProblem: Sensor will not go into portProbable Cause: (1) Insufficient
80095110 rev. 007 01/09TROLL 9500 Operator’s Manual WATER QUALITY ACCESSORIES Basic Sensors Catalog No.pH ...
14 5Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09APPENDIXELECTRONIC DRIFT AND DEVICE RECALIBRATIONThe ele
14 60095110 rev. 007 01/09TROLL 9500 Operator’s Manual g x (L2 - L1)(P2 – P1) x 6.894757R = Note: Density can also be measured using a hy
14 7Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09GLOSSARYABend: Indication in the software interface that
14 80095110 rev. 007 01/09TROLL 9500 Operator’s Manual Cation: Positively charged ion (e.g., NH4+, Na+).Cell constant: A value that describes a cond
1490095110 rev. 007 01/09TROLL 9500 Operator’s Manual Host PC: Desktop computer running Win-Situ, or PDA running Pocket-Situ, connected to the MP TR
1500095110 rev. 007 01/09TROLL 9500 Operator’s Manual dimeter. Comparable to previously reported Formazin Turbidity Unit (FTU), and Jackson Turbidit
1510095110 rev. 007 01/09TROLL 9500 Operator’s Manual Reading: Measurement from a single channel (parameter).Redox potential, see Oxidation-reductio
1520095110 rev. 007 01/09TROLL 9500 Operator’s Manual Surface connector: Multi-function vented nylon connector at the sur-face end of the MP TROLL’s
153Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09IndexAABEND, 37, 39, 143Accessories, 8–9Accuracy specifica
154INDEXTROLL 9500 Operator’s Manual 0095110 rev. 007 01/09Cleaning, 4, 142ammonium sensor, 109chloride sensor, 114conductivity sensor, 77dissolv
90095110 rev. 007 01/09TROLL 9500 Operator’s Manual SECTION 2: COMPONENTS & FEATURESMP TROLL 9500 Maintenance kit (o-rings, lubricant, dust caps
155INDEXTROLL 9500 Operator’s Manual 0095110 rev. 007 01/09saving data, 138setting unit preferences, 136starting from a file or template, 136start
156INDEXTROLL 9500 Operator’s Manual 0095110 rev. 007 01/09Pressure rating, sensors, 51, 65Pressure sensorabout, 46–47absolute vs. gauged, 47fact
15 7INDEXTROLL 9500 Operator’s Manual 0095110 rev. 007 01/09Test Wizard, 35Text files, 45Top of Casing mode, 48Total Dissolved Solids, 77TROLL Co
221 East Lincoln Avenue Ft. Collins, Co. 80524 USA Tel: 970.498.1500 800.446.7488 Fax: 970.498.1598 Declaration of Conformity Manufacturer:
Declaration of Conformity Manufacturer: In-Situ, Inc. 221 East Lincoln Avenue Fort Collins, CO 80524 USA Declares that the following product:
10TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09PRODUCT SPECIFICATIONS7ETTED-ATERIALS 06#TITANIUM6ITON®, acetal, 316L SS /PERATING4EMPER
11TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09pH Type Glass sensing bulb, single-junction electrode, replaceable ceramic junction, refillable
12Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09This section provides a quick overview of the initial step
Copyright © 2001 – 2009 by In-Situ Inc. All rights reserved. Revision history Beta draft September 23, 2005 Initial release, October 28, 2005
130095110 rev. 007 01/09TROLL 9500 Operator’s Manual INSTALL SENSORSBasic water-quality sensors—pH or pH/ORP, polarographic D.O., Conductivity—may a
140095110 rev. 007 01/09TROLL 9500 Operator’s Manual INSTALL THE SOFTWARE WIN-SITU 4 Install Win-Situ 4 from the In-Situ software CD or the In-Situ
15TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09TWIST-LOCK CABLE CONNECTIONSRuggedCable mates to the TROLL 9500 and to Cable Connect TROLL Coms
160095110 rev. 007 01/09TROLL 9500 Operator’s Manual Q:A:What do I do if the MP TROLL 9500 does not appear in the Navigation tree?First try double-c
170095110 rev. 007 01/09TROLL 9500 Operator’s Manual QUICK-CAL THE BASIC WATER-QUALITY SENSORSYou will need: s -042/,,WITH"ASICSENSOR
180095110 rev. 007 01/09TROLL 9500 Operator’s Manual The Quick Calibration Wizard starts. A screen like this is displayed. 7. Insure there is a c
190095110 rev. 007 01/09TROLL 9500 Operator’s Manual s !LTERNATIVELYYOUMAYCLICK!CCEPTTOSTORETHEEARLYVALUES4HEAccept button becomes ava
200095110 rev. 007 01/09TROLL 9500 Operator’s Manual The Wizard displays a final calibration screen like the one below at the end of the Quick Cali
210095110 rev. 007 01/09TROLL 9500 Operator’s Manual 12SECTION 3: GETTING STARTEDSET UP TO LOG DATA (SET UP A TEST)A “test” is a set of instructions
22Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/094 CONTROL SOFTWAREThe interface from a desktop or laptop
0095110 rev. 007 01/09Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual IIIContents1 INTRODUCTION ...
230095110 rev. 007 01/09TROLL 9500 Operator’s Manual SECTION 4: CONTROL SOFTWARENavigation treeActionsInformation paneThe Win-Situ 4 Application Win
240095110 rev. 007 01/09TROLL 9500 Operator’s Manual SOFTWARE FUNCTIONSCONFIGURING THE PORTWhen you start Win-Situ or Pocket-Situ for the first time,
25TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09The Information pane also provides buttons for actions you can perform, such as upgrading the de
26TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09EDITING THE DEVICE PROPERTIESSetting a Device NameThe software recognizes each device on the net
270095110 rev. 007 01/09TROLL 9500 Operator’s Manual CHOOSING MEASUREMENT UNITS AND OTHER PREFERENCESPreferences can be set at any time, and are sav
280095110 rev. 007 01/09TROLL 9500 Operator’s Manual CHOOSING PRESSURE DISPLAY & CONVERSION OPTIONSMeasurements from the optional pressure senso
290095110 rev. 007 01/09TROLL 9500 Operator’s Manual LOGGING DATA (ADDING A TEST)Logging data with the MP TROLL 9500 is called “running a test.” To
300095110 rev. 007 01/09TROLL 9500 Operator’s Manual SECTION 4: CONTROL SOFTWAREEXTRACTING AND VIEWING DATALogged data (“tests”) reside in the MP TR
31Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/095 PROFILINGAll models of the Multi-Parameter TROLL 9500 i
320095110 rev. 007 01/09TROLL 9500 Operator’s Manual s 2EADINGSAREUPDATEDAPPROXIMATELYEVERYSECONDSASTHEProfiler cycles through the availab
0095110 rev. 007 01/09TROLL 9500 Operator’s ManualCONTENTSIVCustomizing the Profiler ...
33TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SECTION 5: PROFILINGClose the Profiler and return to the Parameter displayDisplay a graph of the
340095110 rev. 007 01/09TROLL 9500 Operator’s Manual Q:A:Can I see Dissolved Oxygen readings in mg/L and % saturation at the same time?Yes, so long
35Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09TIP: Defining a test, setting up a test, adding a test—thes
36TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09TEST SETUP PARAMETERSA complete test definition has five parts. The choices available in each SCR
37TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09parameter may be selected for event comparisons, and the “delta” or difference amount is prompt
38TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SECTION 6: LOGGING DATAMeasurement Interval“Delta” valueDefault storageEvent Test SetupEvent ch
39TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09OTHER TEST OPERATIONSThere are six basic test operations:s Add...—programs the instrument to c
40TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09STARTING A MANUAL START TEST1. With the MP TROLL 9500 connected to a host C, select a pending
41TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09DELETING TESTSTests are stored in the device memory until you delete them. The De-lete operatio
42TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09TRANSFERRING FILES FROM A PDA TO A DESKTOP PCIf Win-Situ Sync (or its earlier release, Pocket-S
0095110 rev. 007 01/09TROLL 9500 Operator’s ManualCONTENTSVCalibration ...
43TROLL 9500 Operator’s Manual 0095110 rev. 007 01/092. The Data Folder has a node for data from each device type. TROLL 9500 data will be listed u
44TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09Win-Situ Graphing Controls The Graph button displays data from the first four test channels in g
45TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09Pocket-Situ Graphing Controls Pocket-Situ displays all the test channels in one graph. Time is
46Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/097 MONITORING PRESSURE (WATER LEVEL)Pressure sensorWHY MON
47TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09OPERATING PRINCIPLEA pressure transducer senses changes in pressure, measured in force per squa
48TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09DepthSETTING UP PRESSURE MEASUREMENTS: THE PA-RAMETER WIZARDWin-Situ’s Parameter Wizard can hel
49TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09Liquid Density4HECONVERSIONFROM03)UNITSTOMETERSORFEETOFmUIDREQUIRESseveral conversi
50TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09For DEPTH:1 Select Pressure in Parameters list, select Edit2 Display Mode: Select Depth, clic
51TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09INSTALLATION FOR PRESSURE/LEVEL MEASURE-MENTPosition the instrument below the lowest anticipate
52TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09CORRECTING ABSOLUTE PRESSURE READINGS FOR BAROMETRIC PRESSUREPressure measured with an “absolut
0095110 rev. 007 01/09TROLL 9500 Operator’s ManualCONTENTSVI14 OXIDATION-REDUCTION POTENTIAL ... 97What is ORP? ...
53TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SECTION 7: MONITORING PRESSURE (WATER LEVEL)4. In the Barometric Correction window, select the
54Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/098 MONITORING TEMPERATUREPressure/Turbidity or plugTempera
55TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09ROLE OF THE SENSOR IN CALIBRATION The temperature sensor has a primary function during calibrat
56Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/099 MONITORING BAROMETRIC PRESSUREWHAT IS BAROMETRIC PRESSU
57TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09 Elevation Pressure ft m in. Hg mm Hg PSI Bars -1,000 -304.8 31.02 787.9 15
58Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0910 MONITORING WATER QUALITY: OVERVIEWWHY MONITOR WATER QU
59TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09Q:A:Will a sensor work if installed in the wrong port?Physically, a sensor may be plugged into
60TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09Q:A:What is the difference between NOMINAL and STABLE?To meet the criteria for a valid calibrat
61TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09CALCUPBaseQUICK CALIBRATIONA “Quick Cal” calibrates the Basic sensors simultaneously to achieve
62TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09StirrerPropeller compartmentBattery com-partmentMotor com-partmentEnd cap jl off onEFFE
0095110 rev. 007 01/09TROLL 9500 Operator’s ManualCONTENTSVIISensor Care ...
63TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09MP TROLL 9500 A. with restrictor and nose coneB. with Cal Cup in place of restrictor and nose c
64TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SN32072Files:File NameSN32072 2006-04-17 171530 TURB.htmlSN32072 2006-06-18 102309 DO.htmlSN320
65TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09WHEN TO REPLACE A SENSORAfter a certain amount of use even a complete recalibration will not be
66Multi-Parameter Water Quality TROLL®0095110 rev. 007 01/09TROLL 9500 Operator’s Manual 11 pHWHAT IS pH? The term pH is derived from “p” meaning
67TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09sensor removal tool Eo is related to the potential of the reference electrode, R is the Gas L
68TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09s ANDP(BUFFERSOLUTIONSFORNEUTRALTOACIDICCONDITIONSs ANDP(BUFFERSO
69TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0976105. Select the MP TROLL 9500 in the Navigation tree. The software will automatically det
70TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0912. Remove the Cal Cup, discard the first solution, rinse the Cal Cup and the front end of the i
71TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09sensor removal tools P(READINGSINpure water samples (<100μS/cm conductivity—also known a
72Multi-Parameter Water Quality TROLL®0095110 rev. 007 01/09TROLL 9500 Operator’s Manual 12 CONDUCTIVITYUltra-pure distilled water 0.05 μS/cmDist
1Multi-Parameter Water Quality TROLL®TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09GENERAL DESCRIPTION OF THE MP TROLL 9500Your new Multi-Para
73TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09Type Operating Range Cell Constant RangeLow 3 to 50,000 μS/cm 0.33 – 0.39 cm-1 High 70 to
74TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09CONDUCTIVITY QUICK CALThe procedure to Quick Cal the conductivity sensor (a 1-point calibra-tio
75TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0910. In the next screen, select Run to begin the stabilization. The display will continuously u
76TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09USING A CONDUCTIVITY METER AS A SECONDARY STANDARDIf a conductivity meter is available for comp
77TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09Total Dissolved Solids (TDS)A factor of 0.65 x specific conductance is used to estimate Total Di
78Multi-Parameter Water Quality TROLL®0095110 rev. 007 01/09TROLL 9500 Operator’s Manual 13 DISSOLVED OXYGEN$/½#ATMPPM#HLORINE
79TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09THE POLAROGRAPHIC DISSOLVED OXYGEN SENSORTHEORY OF OPERATIONThe sensor is a Clark-type polarogr
80TROLL 9500 Operator’s Manual 0095110 rev. 007 01/094. Thread the membrane module to the D.O. sensor.Be sure the membrane does not leak. You shoul
81TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09depleted solution such as sodium sulfite or nitrogen-saturated water. A 2-point calibration is r
82TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SECTION 13: DISSOLVED OXYGEN—POLAROGRAPHICPreparation for 100% D.O. Calibration in WaterThe bub
20095110 rev. 007 01/09TROLL 9500 Operator’s Manual CONVENTIONSThroughout this operator’s manual you will see the following symbols.The check mark h
83TROLL 9500 Operator’s Manual 0095110 rev. 007 01/097. If not already connected, connect the MP TROLL 9500 to a PC, launch the software, and “find”
84TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0916. Click Next to continue. A screen like this will be displayed:17. Select Run to begin stabil
85TROLL 9500 Operator’s Manual 0095110 rev. 007 01/0920. When readings have stabilized (or you Accept the Nominal value), the final screen is display
86TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09USAGE RECOMMENDATIONS AND CAUTIONSThe amount of oxygen that can be dissolved in water decreases
87TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SENSOR CARE AND HANDLINGThe D.O. sensor kit includes the following items for routine mainte-nan
88TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09SECTION 13: DISSOLVED OXYGEN—POLAROGRAPHICLong-Term StorageRemove the membrane module, rinse wi
89TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09THE RDO OPTICAL DISSOLVED OXYGEN SENSORThe In-Situ RDO ® optical dissolved oxygen sensor measur
90TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09cone from the sensor and adapter assembly. Nose coneRestrictorAdapterRDO sensor The soft Cal Cu
91TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09C 0RESSUNTILTHEINSTRUMENTSSENSORBLOCKSURFACEISmUSHWITHthe adapter plate surface. A
92TROLL 9500 Operator’s Manual 0095110 rev. 007 01/09 Allow 5-10 minutes for temperature and oxygen equilibration.TIP: The equilibration time depen