Instruction SheetCommunication Device KitPart number 0081100Kit Contentsl Communication devicel Software CDl 9 V batteryl You can also purchase a cali
800-446-7488 10 www.in-situ.com11. Enter the barometric pressure or elevation at which the instrument will be deployed,if necessary.12. Click Next.13.
800-446-7488 11 www.in-situ.comcalibration. User calibration is recommended only if you must conform to a standardoperating procedure or if the conduc
800-446-7488 12 www.in-situ.com7. Select the appropriate calibration standard from the drop-down list. If you select“User Defined,” enter the value of
800-446-7488 13 www.in-situ.compH/ORP Sensor CalibrationFill the calibration cup to the fill line with approximately 130 mL of the desired pH orORP ca
800-446-7488 14 www.in-situ.com14. Once calibration is complete, remove the instrument from the calibration cup andrinse both thoroughly with clean wa
800-446-7488 15 www.in-situ.comDiagnostics and Factory ResetThe Probe Info Tab displays the software application version, probe type, and probeserial
800-446-7488 2 www.in-situ.comappropriate wire in the location specified by the diagram.2. Attach the USB connector to a USB port on the computer.3. W
800-446-7488 3 www.in-situ.comView Live ReadingsWhen the instrument is connected to the software, a snap shot of data appears onscreen. If you want li
800-446-7488 4 www.in-situ.comAdjust Communications OutputsThe Communication Tab allows you to configure your instrument to communicate viaModbus or S
800-446-7488 5 www.in-situ.comSDI-12 SetupSDI-12 setup allows you to set the instrument address, select the parameters youintend to log, and select th
800-446-7488 6 www.in-situ.comSensor Setup TabThe Sensor Setup Tab allows you to configure and calibrate sensors, assign thesentinel value, and set un
800-446-7488 7 www.in-situ.comRDO SensorScreen Element PurposeEnable automatic salinitycorrection from the conductivitysensor (checkbox)Select the che
800-446-7488 8 www.in-situ.comScreen Element PurposeSentinel value(text field)Displays the current sentinel value. The sentinel valueis the number tha
800-446-7488 9 www.in-situ.com5. Wait 5 to 10 minutes for temperature stabilization prior to calibration.Note: Do not leave the instrument in the cali