Résumé du contenu

Page 1 - Operator's Manual

Operator's ManualCon TROLL®PROInformation subject to change without notice. In-Situ, In-Situ logo, Baro Merge, BaroTROLL, HERMIT, iSitu, Pocket-S

Page 2

800-446-7488 10 www.in-situ.coml ON—The back-light stays on continuously.l Interval—The back-light stays on for the specified interval after the last

Page 3 - Table of Contents

800-446-7488 11 www.in-situ.comOn-Screen Icon DescriptionsError with parameterorRDO PRO (RDO Classic Sensor Cap) has expired.Error with parameterCalib

Page 4 - 5 Aqua TROLL 400 Instrument

800-446-7488 12 www.in-situ.comIMPORTANT: Any change to the configuration of the parameters(add probe, delete probe, enable/disable parameters, change

Page 5 - 9 Troubleshooting

800-446-7488 13 www.in-situ.com3. When pairing between the controller and the Bluetooth device is successful thecontroller screen will show “Pairing S

Page 6 - Getting Started

800-446-7488 14 www.in-situ.comConfiguring ProbesAdding a Probe After Initial Configuration or Deletion1. From the main screen, select Menu, Settings,

Page 7 - Initial Configuration

800-446-7488 15 www.in-situ.comSetting Parameter Units1. From the main screen, select Menu, Settings, Probes.2. Select a probe.3. Select parameters.4.

Page 8 - Using the View Function

800-446-7488 16 www.in-situ.comRDO PRO or RDO PRO-X ProbeConfiguring RDO PRO or RDO PRO-X OptionsThe RDO PRO and the RDO PRO-X probes are both compati

Page 9 - Display (Screen) Appearance

800-446-7488 17 www.in-situ.coml Concentration—Choose this option if you want to calibrate the concentrationvalues to a concentration obtained using a

Page 10

800-446-7488 18 www.in-situ.comFigure 3.3 Beginning the 1-point calibration; accepting or canceling the calibration.Controller 2-Point Calibration (10

Page 11 - Setting the Log Interval

800-446-7488 19 www.in-situ.com6. You will see the Calibration Beginning icon next to the % Sat value. When thecalibration is complete, choose Stable

Page 12

800-446-7488 2 www.in-situ.comCopyright © 2009-2013 by In-Situ Inc. All rights reserved.This document contains proprietary information which is protec

Page 13

800-446-7488 20 www.in-situ.comAqua TROLL 100 and 200 InstrumentsConfiguring Aqua TROLL 100 and 200 OptionsConductivity/Alpha Coefficient1. From the m

Page 14 - Configuring Probes

800-446-7488 21 www.in-situ.comLevel/Specific Gravity (Aqua TROLL 200 only)1. From the main screen, select Menu, Settings, Probes.2. Select the Aqua T

Page 15 - Setting Parameter Units

800-446-7488 22 www.in-situ.com4. You will see the Calibration Beginning icon next to the μS/cm value. When thecalibration reaches stability, choose S

Page 16 - RDO PRO or RDO PRO-X Probe

800-446-7488 23 www.in-situ.comAqua TROLL 400 InstrumentConfiguring the Aqua TROLL 400 Sensor OptionsBarometer1. From the main screen, select Menu, Se

Page 17

800-446-7488 24 www.in-situ.com1. From the main screen, select Menu, Settings, Probes.2. Select the Aqua TROLL 400 Instrument.3. Select Options.4. Sel

Page 18

800-446-7488 25 www.in-situ.com4. Select Pressure, Specific Gravity.5. Select “Fixed” or “Auto.”l If you select “Fixed,” use the arrow keys to enter a

Page 19 - Calibration Hints

800-446-7488 26 www.in-situ.coml Cal. Report—Choose this option to view the date of the last calibration. Press theMore button to view the Slope and O

Page 20

800-446-7488 27 www.in-situ.com7. You will see the “Calibration Beginning” icon next to the mg/L reading. When thecalibration is complete, select Stab

Page 21

800-446-7488 28 www.in-situ.com4. You will see the Calibration Beginning icon next to the % Sat value. When thecalibration is complete, choose Stable

Page 22

800-446-7488 29 www.in-situ.com7. If you select Stable, the Standard Value screen appears. The instrument auto-detects the standard solution in which

Page 23 - Aqua TROLL 400 Instrument

800-446-7488 3 www.in-situ.comTable of Contents1 Getting Started6Probe Installation Requirements 6Navigating Through the Menus 6Menu Diagram 7Initial

Page 24

800-446-7488 30 www.in-situ.com9. When you select Stable, the Standard Value screen appears. The instrument auto-detects the standard buffer in which

Page 25

800-446-7488 31 www.in-situ.comCal. Report1. From the main screen, select Menu, Calibrate.2. Select the Aqua TROLL 400 Instrument.3. Select the approp

Page 26

800-446-7488 32 www.in-situ.comLevel TROLL InstrumentConfiguring Level TROLL OptionsSetting the Level Mode1. From the main screen, select Menu, Settin

Page 27

800-446-7488 33 www.in-situ.comCalibration HintsNominal Stability vs. Full StabilityTo meet the criteria for a valid calibration point, the change (de

Page 28

800-446-7488 34 www.in-situ.comCurrent Loops and RelaysConfigure the Current LoopsIf you have wired 1 or 2 current loop outputs as described in the Co

Page 29 - Calibrating Pressure/Level

800-446-7488 35 www.in-situ.comAlarms Setpoints and DeadbandsHigh Alarm Model The relay is turned on if the current reading is equal to or greater tha

Page 30

800-446-7488 36 www.in-situ.comLogged DataCon TROLL PRO Models AC-L and DC-L have logging capabilities. Parameters arelogged according to a set value

Page 31

800-446-7488 37 www.in-situ.com3. Set the Con TROLL PRO in the Enable Pairing mode. From the main screen selectMenu, Settings, Communication, Bluetoot

Page 32 - Level TROLL Instrument

800-446-7488 38 www.in-situ.com7. Click the Finish button.8. Open the Bluetooth Devices utility.9. Select the Options tab. Select the check boxes show

Page 33

800-446-7488 39 www.in-situ.com10. Click the COM Ports tab. Note the Port number of the Outgoing port (in this example,Port 17). You will need to sele

Page 34 - Current Loops and Relays

800-446-7488 4 www.in-situ.comNominal Stability vs. Full Stability 19Factory Recalibration 194 Aqua TROLL 100 and 200 Instruments20Configuring Aqua TR

Page 35

800-446-7488 40 www.in-situ.com3. Click the "New Connection" button to start the connection Wizard. Select the"Express Mode" optio

Page 36 - Logged Data

800-446-7488 41 www.in-situ.com6. Each time you connect, you will have to enter the passkey. The passkey for the ConTROLL PRO is by default its serial

Page 37

800-446-7488 42 www.in-situ.com2. After wiring the stripped-and-tinned Rugged Cable, connect either a cable connectUSB or RS232 TROLL Com communicatio

Page 38 - 7. Click the Finish button

800-446-7488 43 www.in-situ.com4. Click the Connect button. The Home screen appears. It displays the ambienttemperature and barometric pressure measur

Page 39

800-446-7488 44 www.in-situ.com7. Click the My Data tab shown in the figure above. You can view your data in the rightwindow or download it as a *.csv

Page 40

800-446-7488 45 www.in-situ.comDownloading Data using Win-Situ Mobile and a RuggedReader Handheld PCConfiguring the Bluetooth Utility on the RuggedRea

Page 41

800-446-7488 46 www.in-situ.com5. Click Done, then select the Com Ports tab. Tap New Outgoing Port.6. Select your Con TROLL PRO device. Tap Next.7. Ma

Page 42

800-446-7488 47 www.in-situ.comConnecting the Con TROLL PRO to Win-Situ Mobile Software1. Tap Start, then Win-Situ Mobile.2. Tap File, then Comm Setti

Page 43

800-446-7488 48 www.in-situ.com4. A series of screens will appear. Move through them by tapping the forward arrowbutton.l Make sure that the Port Numb

Page 44

800-446-7488 49 www.in-situ.com7. Tap the Con TROLL PRO log to highlight it. Then tap the Download button. Selectthe how much data to download: All, N

Page 45

800-446-7488 5 www.in-situ.comDefaults 316 Level TROLL Instrument32Configuring Level TROLL Options 32Setting the Level Mode 32Entering the Specific Gr

Page 46

800-446-7488 50 www.in-situ.com10. To export the data to csv format, tap File, then Options. Tap Export All Now, then theCheck Mark button to export a

Page 47

800-446-7488 51 www.in-situ.comTroubleshootingDiagnosticsCon TROLL PRO diagnostics allow you to find serial numbers, firmware and hardwareversions, ca

Page 48

800-446-7488 52 www.in-situ.comFinding Calibration Data and Constants1. Select Menu, Diagnostics.2. Select the probe, then Enter.3. Scroll down to the

Page 49

800-446-7488 53 www.in-situ.comTroubleshooting Power IssuesDangerOnly properly trained and qualified personnel should troubleshoot the powerconnection

Page 50 - Bluetooth Special Hints

800-446-7488 54 www.in-situ.comTroubleshooting Probe Connection Issues

Page 51 - Troubleshooting

800-446-7488 55 www.in-situ.comTroubleshooting Bluetooth Communications Issues

Page 52

800-446-7488 6 www.in-situ.comGetting StartedProbe Installation RequirementsBefore making any wiring connections between the probes and the module, ma

Page 53 - Troubleshooting Power Issues

800-446-7488 7 www.in-situ.comMenu DiagramFigure 1.2 ConTROLL PRO Menu DiagramInitial ConfigurationEnter the following information upon first use or a

Page 54

800-446-7488 8 www.in-situ.comRegion UTC Value (± numeric value)Australia Australian CapitalTerritory+10 (Daylight saving time +11)Australia South +9.

Page 55

800-446-7488 9 www.in-situ.comFigure 1.4 Probe A, probe B and probe CProbe A parameters are labeled with A. Probe B parameters arelabeled with B, and