Input/Output Connections 25
CONNEXIONS POUR LE RACORD DES SONDES
Danger
Débrancher toute alimentation à l'appareil avant de
connecter les fils.
1. Disconnect power to the instrument.
2. Use a Phillips screwdriver to remove the controller lid.
Note: If you want to use
RuggedCable
®
with
twist-lock connectors,
order a short length of
stripped and tinned
RuggedCable with a
male connector. This
can be wired into the
controller, then
connected to existing
twist-lock cable
connectors.
3. Connect the stripped and tinned wires from the RuggedCable to the
terminal strip as follows:
4. Clip or cap the unused wires (brown and white).
5. Replace the controller lid and tighten screws.
6. Reconnect power to the device.
FIGURE 21. Earth ground
5.3 CURRENT LOOP OUTPUTS
Danger
Make sure that power to the instrument is disconnected
before making any wiring connections.
Danger
The maximum voltage that can be applied across the loop
terminals is 36 V.
SORTIES DE COURANT DE BOUCLE
Danger
Débrancher toute alimentation à l'appareil avant de
connecter les fils.
Danger
La tension maximum à travers la boucle ne doit pas
excéder 36 V.
1. Disconnect power to the instrument.
2. Use a Phillips screwdriver to remove the controller lid.
GND Black
485A- Green
485B+ Blue
POWER Red
Green Earth Ground screw (Figure 21) Shielded wire (silver color)
Earth ground screws for
sensor silver wires