Specifications 31
Chapter 8 Specifications
8.1 OVERVIEW
The Con TROLL PRO controller is a process sensor controller with an easy-
to-use interface that supports two sensors. AC and DC versions are
available:
• The DC version is low-power consumption data logger with local display and
communication gateway for In-Situ groundwater or surface water probes and
sensors.
• The AC version is a line-powered process controller with optional logging
functions, plus local display and communication gateway features for In-Situ
process instruments and sensors.
8.2 SPECIFICATIONS
Component Description
Microprocessor-controlled measuring unit with measured value display,
temperature display, and menu-driven system.
Controller Operating Temperature –20 to 70 °C (–4 to 158 °F); 95% relative humidity, non-condensing
Controller Storage Temperature –40 to 85 °C (–40 to 185 °F); 95% relative humidity, non-condensing
Logging Memory 4 MB
Barometric Pressure Accuracy ±3 mbar max.
Barometric Pressure Resolution 0.01 mbar
Ambient Temperature Accuracy ±2° C maximum
Ambient Temperature Resolution 0.1° C
Enclosure NEMA 4X/IP67
Power Requirements
Model AC and Model AC-L: 100–240 V
~, 0.15 A, 50–60 Hz
Model DC-L: 9–36 V , 0.2 A max
Outputs (2) 4-20 mA isolated current loop
Power Draw
Con TROLL Pro + 1 RDO Pro idle (not measuring, display on): 10 mA typical
Con TROLL Pro + 1 RDO Pro active (measuring, display on): 20 mA typical
Con TROLL Pro + 1 RDO Pro sleep (not measuring, display off): 320 µA typical
Con TROLL Pro only, sleep (no probe connected, display off): 70 µA typical
Relays
(2) low voltage (<50 V) max at 2A
(2) high voltage (>50 V) (AC-powered model only) 264 VAC max at 5A
Controller Dimensions 6.3 X 6.3 X 3.56 in (16 X 16 X 9.04 cm)
Controller Weight
AC and AC-L Models: 3.0 lb
DC-L Model: 2.4 lb
Certifications
CE approved (with RDO
®
PRO Optical Dissolved Oxygen sensor)
Listed for use in general locations to UL and CSA safety standards by
ETL (with RDO PRO Optical Dissolved Oxygen sensor)
Commentaires sur ces manuels